Selection knob

• Press to reach the set-up menu of the function

• Turn to choose the parameter you want to modify

• Press to select the parameter you want to modify

• Turn to change the set value of the parameter

• Press to confirm change of value and to reach further

parameters, or to leave the set-up menu

Notes on connecting and operating the controller

Valves and/or vacuum pumps are necessary to operate the control-

ler.

Without those components the controller can be used only as vacuum

measurement device.

When used as measurement device, the controller can switch a possibly

connected coolant valve.

The CVC 3000 is equipped with an internal capacitive pressure trans-

ducer with ceramic diaphragm. It measures the actual pressure indepen-

dently of the gas type, and with reference to the vacuum, i.e., absolute.

➨

Connect the measurement connection of the controller via hose or

PTFE tubing to the vacuum application, or connect an external pres-

sure transducer 3000 series. The cross-section of the tubing should

be as large as possible.

Assembling the country-specific power plug

+

The wall power supply is delivered with

power plugs for Europe, UK, US and Aus-

tralia.

+

Press the locking key at the wall power sup-

ply to remove and to replace the power plug

with your country-specific plug.

+

Assemble the suitable power plug to the

power supply and lock.