Assembling the country-specific power plug, Notes on connecting and operating the controller – VACUUBRAND CVC 3000 User Manual

Page 26

Advertising
background image

page 26 of 80

Selection knob

• Press to reach the set-up menu of the function

• Turn to choose the parameter you want to modify

• Press to select the parameter you want to modify

• Turn to change the set value of the parameter

• Press to confirm change of value and to reach further

parameters, or to leave the set-up menu

Notes on connecting and operating the controller

Valves and/or vacuum pumps are necessary to operate the control-

ler.

Without those components the controller can be used only as vacuum

measurement device.

When used as measurement device, the controller can switch a possibly

connected coolant valve.

The CVC 3000 is equipped with an internal capacitive pressure trans-

ducer with ceramic diaphragm. It measures the actual pressure indepen-

dently of the gas type, and with reference to the vacuum, i.e., absolute.

Connect the measurement connection of the controller via hose or

PTFE tubing to the vacuum application, or connect an external pres-

sure transducer 3000 series. The cross-section of the tubing should

be as large as possible.

Assembling the country-specific power plug

+

The wall power supply is delivered with

power plugs for Europe, UK, US and Aus-

tralia.

+

Press the locking key at the wall power sup-

ply to remove and to replace the power plug

with your country-specific plug.

+

Assemble the suitable power plug to the

power supply and lock.

Advertising