Auto mode (only nt vario / vario-b pumps), Auto mode, Only nt vario / vario-b pumps) – VACUUBRAND CVC 3000 User Manual

Page 39

Advertising
background image

page 39 of 80

Auto mode

(only NT VARIO / VARIO-B pumps)

Preselections
+

Use the selection knob to set the parameters.

+

Sensitivity: The ”Sensitivity” of the control determines the control speed.

The ”Sensitivity” is adjustable to ”high”, ”normal” or ”low”.

High sensitivity leads to a reduced pump-down speed, e.g., for small

amounts of solvents or foaming processes. Use a low sensitivity for ”non-

critical” processes to reduce process time. Usually setting the ”Sensitiv-

ity” to ”normal” is appropriate.

+

Speed: The maximum motor speed can be preselected (to limit the

pumping speed).

The ”Speed” is adjustable in a range of 1-100% or can be set to ”HI”.

The selection ”HI” (recommended) causes an automatic limitation of

the adaptation depending on parameters determined during the pro-

cess. Only in case of very sensitive processes is it advisable to reduce

the maximum speed.

+

Minimum: If the vacuum value reaches ”Minimum” the process is

stopped. Use ”Minimum” for automatic process termination.

The ”Minimum” is adjustable in a range of 2-795 Torr (2 - 1060 mbar),

or can be set to ”Auto” or to ”Off”.

When selecting ”Auto” the controller switches off automatically at the

end of the process (boiling plateau). This is recommended only for

large amounts of solvents.

+

Delay: Delay” specifies the time the coolant valve remains open after

process stop.

The ”Delay” is adjustable in a range of 1-300 minutes or can be set to

”Off” (”Off” means that the coolant valve closes immediately when the

process stops.).

Control of a NT VARIO / VARIO-B pump in ”Auto mode”:

Automatic detection of boiling points, and automatic adjustment

of determined boiling points as process parameters change.