Mbar – VACUUBRAND CVC 3000 User Manual

Page 55

Advertising
background image

page 55 of 80

Note: The active pressure transducer

(VSP 3000) is not displayed in the sta-

tus line of the controller display!

When operating the CVC 3000 with the

external VSP 3000 pressure transduc-

er,

vacuum control is not possible in

the

”Auto mode” and ” VACUULAN”

functions.

1.0

x

10

3

mbar

”Pump down” function:

Minimum” is adjustable in a range of 1x10

-3

Torr to 7.5x10

2

Torr (1x10

-3

mbar to 1x10

3

mbar) or can be set to ”Off”.

”VAC control” function:

Set vacuum” (factory set 7.5x10

2

Torr), ”Hysteresis” and ”Maximum” are

adjustable in a range of 7.5x10

2

Torr to 1x10

-3

Torr (1x10

3

mbar to 1x10

-3

mbar).

Set vacuum in Torr

1-9 x 10

1

1-9 x 10

0

1-9 x 10

-1

1-9 x 10

-2

1-9 x 10

-3

Hysteresis in Torr

(suggested values)

1 x 10

1

1 x 10

0

1 x 10

-1

1 x 10

-2

1 x 10

-3

Adjustment of the set vacuum during vacuum control:

Fine tuning:

The set vacuum can be fine-adjusted by turning the selection knob while

process is running.
+

Turn the selection knob.

+

A full turn causes a change of the set vacuum of 1.2 decades.

+

Turning the knob one detent causes a change of the set vacuum of

1/10 decade.

Serial Interface:

Command IN_PV: The vacuum value is given as X.XXE±X.

Command OUT_SP or IN_SP: The vacuum value is given as XE±X.