Adjustment at a reference pressure

Instead of adjustment under vacuum to a pres-

sure < 0.1 Torr (mbar), adjustment to a precise-

ly known reference pressure within the range

of 0..... 15 Torr (20 mbar) is possible.

Evacuate the measurement connection of the CVC 3000 or in case

that of a connected optional external gauge head VSK 3000 to a pres-

sure within 0 ..... 15 Torr (0.....20 mbar).

In ”Configuration” menu, select program ”Adjustment” at the controller.

+

The reading is automatically adjusted to ”zero”.

Use the selection knob to adjust the display to the reference pressure

at the vacuum line within the range of 0 ..... 15 Torr (0.....20 mbar).

Press the selection knob to confirm.

0...15 Torr

Note: The accuracy of the value of the reference pressure will directly

affect the accuracy of the adjustment. If the nominal ultimate vacuum

of a diaphragm pump is used as reference vacuum, the accuracy of the

controller might be doubtful. The diaphragm pump may not achieve the

specified value (due to condensate, poor condition, failure of valves or

diaphragm, leaks).

Gauge head VSP 3000 (accessory)

Read the manual of the gauge head VSP 3000 carefully before in-

stalling or operating the equipment. Observe the instructions con-

tained in that manual.

The display and measuring range of the CVC 3000 with gauge head VSP

3000 is from 7.5 x 10

2

Torr to 1 x 10

-3

Torr (1 x 10

3

mbar to 1 x 10

-3

mbar).

The pressure reading is given in exponential notation.

+

Attention: Maximum pressure reading: 7.5x10

2

Torr

(1x10

3

mbar). Pressure values above 7.5x10

2

Torr

(1x10

3

mbar) are not displayed.

Risk of unnoticed

overpressure. Risk of bursting!