VACUUBRAND CVC 3000 User Manual

Page 35

Advertising
background image

page 35 of 80

The screen-shot shows the factory-set values.

When selecting ”Graphic” the display shows a pressure vs. time curve.

The timeline in the diagram adapts automatically to the process time.
+

Press the selection knob twice to return to the standard display.

10 mbar

00:03:50

Pump down

Pump down
Speed

HI

Minimum

Off

Delay

Off

Duration

Off

- - - - - - Graphic - - - - - -

- - - - - - - Back - - - - - - -

760 Torr

Temporary switching from ”Pump down” to ”Vac control” or ”Auto

mode” (only if control is running):
+

Press ”MODE” key. The controller switches to ”Vac control” function,

the current vacuum is used as set value.

+

Pressing ”MODE” key again switches to ”Auto mode” function. The

controller adapts the boiling pressure starting from the current vacuum

level.

+

The preset function of the controller does not change due to this tem-

porary switching. When pressing ”STOP” key, the controller is set

again to the ”Pump down” function.

Adapting the pumping speed during pump down:
+

Press the selection knob and turn.

+

Turning the knob to the left reduces the speed.

+

Turning the knob to the right increases the speed.

be set to ”Off”. ”Off” indicates that no endpoint for pump down is deter-

mined.

If a ”Duration” is preset, the controller switches off the pump when the

preset process time is reached, even if a preset ”Minimum” is still not

reached.

+

If neither ”Minimum” nor ”Duration” is preset, process control has to be

stopped by pressing the ”STOP” key.