Connecting several external gauge heads, Adjustment under vacuum – VACUUBRAND CVC 3000 User Manual

Page 57

Advertising
background image

page 57 of 80

Adjustment under vacuum

An adjustment under vacuum is only possible

in the lower pressure range.

Evacuate the measurement connection of the VSP 3000 pressure

transducer preferably below 7.5x10

-4

Torr (1x10

-3

mbar), e.g., by ap-

plying a high vacuum pump. Allow the pressure transducer a period of

minimum 20 minutes to preheat at this pressure.

In ”Configuration” menu, select program ”Adjustment” at the controller.

+

The reading is automatically adjusted to ”zero”. This value is fix and

can not be changed.

Press the selection knob to confirm.

0 Torr

Connecting several external gauge heads

It is possible to connect several external gauge heads simultaneously to

the CVC 3000 controller. Maximum number of gauge heads of the same

type (VSK 3000 or VSP 3000) connected to one CVC 3000: four. That is

a total of eight external gauge heads at maximum. Up to four additional

gauge heads VSK 3000 may be connected as reference if configured

accordingly (see ”Measuring differential pressure”). Only the readings of

one gauge head are displayed at a time.


+

The VACUU•BUS address of the active gauge head is not shown in

the status line. The readings of all gauge heads are shown in the menu

”Sensors”.

+

Each gauge head needs an individual address (VSK 1 to VSK 4 /

VSP 1 to VSP 4) to communicate with the CVC 3000 Controller via

VACUU•BUS. All gauge heads are shipped with their VACUU•BUS

address factory-set to VSK 1 or VSP 1.

+

Use your CVC 3000 Controller to reconfigure the gauge head (e.g.,

changing the VACUU•BUS address from VSK 1 to VSK 2), see section

”Reconfiguring the gauge head”.