Ambient conditions, Notice – VACUUBRAND CVC 3000 E User Manual

Page 15

Advertising
background image

page 15 of 75

to be sure that the power source and the equipment

match in voltage, phase, and frequency.

• When working with residues, aggressive or condens-

able media, install a gas washing bottle if necessary.

Using a wall power supply, assemble and lock the suit-

able power plug (included in shipment) to the power sup-

ply prior to use.

The controller is equipped with a short-circuit-proof, wide-

range power supply with an integrated overload protec-

tion.

Avoid high heat supply (e.g., due to hot process gases).

Position the controller CVC 3000 E and its vacuum line in

such a way that condensate cannot flow towards the pres-

sure transducer.

In case connect inert gas at the venting connection.

Allow the equipment to equilibrate to ambient temperature

if you bring it from cold environment into a room prior to

operation. Notice if there is water condensation on cold

surfaces.

Comply with all

applicable and relevant safety require-

ments (regulations and guidelines). Implement the re-

quired actions and adopt suitable safety measures.

Ambient conditions

Do not reach for this product if it has fallen into liquid.

There is a risk of deadly electrical shock. Unplug the

system immediately.

+

Do not use this product in an area where it can fall or

be pulled into water or other liquids.

• Adopt suitable measures in case of differences from

recommended conditions, e.g., using the equipment

NOTICE

Advertising