Application examples, Vacuum for filtration and suction, Assembly of a vacuum system – VACUUBRAND CVC 3000 E User Manual

Page 42

Advertising
background image

page 42 of 75

This screen-shot shows the factory-set values.

VACUULAN

40 mbar

00:13:00

When selecting ”Graphic” the display shows a pressure vs. time curve.

The timeline in the diagram adapts automatically to the process time.
+

Press the selection knob twice to return to the standard display.

Assembly of a vacuum system

+

Assemble vacuum connection lines between CVC 3000 E controller,

vacuum pump (diaphragm pump with in-line valve or Vacuum-Man-

agement-System) and vacuum application.

+

Assemble electrical connections.

+

Connect coolant if necessary.

Vacuum for filtration and suction

+

Select ”

Pump down” function.

+

Set ”Minimum” to a value which provides adequate suction but does

not lead to evaporation of the solvent.

+

Start process by pressing ”START/STOP” key.

alternatively:

+

Select ”

Vac control” function.

+

Set ”Set vacuum” (and ”Hysteresis”, if necessary) to a value which

Application examples

VACUULAN
Set vacuum 19 Torr

Switch on 150 Torr

Delay

15

min

- - - - - - Graphic - - - - - -

- - - - - - - Back - - - - - - -

Advertising