page 56 of 75

Interface parameters

The CVC 3000 E controller is equipped with a serial interface (RS 232C,

nine-pin Sub-D-plug).
+

Plug-in or remove the cable (cable RS 232C) from the interface only if

the equipment is switched off.

+

The interface is

not electrically isolated from the measuring circuit.

+

For optimal electromagnetic compatibility assemble an interface filter

(cat. no.: 638235).

The controller is fully operable via the serial interface. Measuring results,

preselections and the status of the controller can be read at any time.

The factory-set read and write commands are completely compatible

with the VACUUBRAND CVC 2000 controller (see sections ”Read / Write

commands CVC 2000”). An extended instruction set is available using the

command ”CVC 3” (see sections ”Read / Write commands CVC 3000”).

Setting of the interface

Set the interface parameters directly at the CVC 3000 E controller. The

factory set values are underlined.

Edit and confirm the interface parameters in the ”Configuration” menu in

”RS-232” submenu using the selection knob.

➨

Baud: 2400, 4800, 9600 or 19200

➨

Parity: 8-N-1, 7-O-1 or 7-E-1

➨

Handshake: Off, Xon-Xoff or RTS-CTS

➨

Remote: On or Off

➨

Timeout: Sending 1s, receiving 10s.

Readjustment of the controller CVC 3000 E

See section “Readjustment”, pg. 47.

➨

Drain the solvent and dispose of in accordance with regulations. Re-

peat cleaning if necessary.

➨

Rinse the measurement chamber several times with alcohol in order

to remove all solvent residues.

➨

Allow the pressure transducer to dry.

➨

Readjust the pressure transducer if necessary.