Fault

Possible cause

Remedy

❑ Warning triangle

and black valve

symbol are flash-

ing, two blips*.

➨ External venting valve

removed or defective?

✔ Connect valve or replace

with a new one or recon-

figure without valve.

❑ Warning triangle

and valve symbol

are flashing, three

blips*.

➨ NT VARIO / VARIO-B

pump and in-line valve

connected?

✔ Disconnect in-line valve;

switch controller off/on to

reconfigure.

➨ In-line valve removed

or defective?

✔ Check connection cable

of the valve; or use new

valve or reconfigure with-

out valve.

❑ Warning triangle

and coolant valve

symbol are flash-

ing, four blips*.

➨ Coolant valve removed

or defective?

✔ Check connection cable

of the valve; or use new

valve or reconfigure with-

out valve.

❑ No digital pressure

reading, warning

triangle is flashing,

five or seven blips*.

➨ Five blips:

External pressure

transducer defective or

removed?

✔ Plug in external pres-

sure transducer, or use a

new one, or reconfigure

without external pressure

transducer.

➨ Seven blips:

Internal pressure trans-

ducer defective?

✔ Contact local distributor.

❑ Warning triangle

and pump symbol

are flashing, six

blips*.

➨ NT VARIO / VARIO-

B pump and VMS**

(Vacuum Management

System) connected?

✔ Remove VMS. Restart

controller.

➨ Fault at the NT VARIO

/ VARIO-B pump or at

VMS?

✔ Check pump or VMS,

restart controller.

➨ Connecting cable from

pump or VMS removed?

✔ Check cable connections.

❑ Clock symbol is

flashing.

➨ Preselected process

time is over?

✔ Confirm by pressing

START/STOP key.

❑ Internal venting

valve does not re-

spond, valve sym-

bol is not displayed.

➨ External pressure

transducer connected

and active?

✔ Select internal pressure

transducer or connect ex-

ternal venting valve.

❑ Venting valve does

not respond, valve

symbol is dis-

played.

➨ Venting valve contami-

nated?

✔ Clean valve.

Advertising