Readjustment, Notice – VACUUBRAND CVC 3000 E User Manual

Page 47

Advertising
background image

page 47 of 75

The vacuum gauge was adjusted using factory standards,

which are traceable through regular calibration in an ac-

credited laboratory (DAkkS calibration laboratory) to the

German national pressure standard. Depending on the

process and/or accuracy requirements, check the adjust-

ment and readjust if necessary. For readjustment, the

device has to be adjusted both at atmospheric pressure

as well as under vacuum but only if the reference pres-

sures are known with certainty. The adjustment mode can

be activated only if the process control is inactive. Press

”START/STOP” key, if necessary. In the range between 15

to 525 Torr (20 to 700 mbar) no adjustment is possible;

---- Torr is displayed.

NOTICE

Readjustment

Adjustment at atmospheric pressure

An adjustment at atmospheric pressure is only

possible if the pressure is higher than 525 Torr

(700 mbar).

➨

Vent the measurement connection of the CVC 3000 E or in case that

of a connected optional external gauge head VSK 3000. Make sure

that the pressure transducer is at atmospheric pressure.

➨

In ”Configuration” menu, select program ”Adjustment” at the controller.

➨

Use the selection knob to adjust the reading to the current atmospheric

pressure.

➨

Press the selection knob to confirm.

760 Torr

Note: To determine the actual atmospheric pressure, use an accurate

barometer or get accurate reading from the weather service, or a nearby

airport or other reliable source (taking into account the difference in alti-

tude between the source and the laboratory).