Readjustment, Notice – VACUUBRAND CVC 3000 E User Manual

Page 47

Advertising
background image

page 47 of 75

The vacuum gauge was adjusted using factory standards,

which are traceable through regular calibration in an ac-

credited laboratory (DAkkS calibration laboratory) to the

German national pressure standard. Depending on the

process and/or accuracy requirements, check the adjust-

ment and readjust if necessary. For readjustment, the

device has to be adjusted both at atmospheric pressure

as well as under vacuum but only if the reference pres-

sures are known with certainty. The adjustment mode can

be activated only if the process control is inactive. Press

”START/STOP” key, if necessary. In the range between 15

to 525 Torr (20 to 700 mbar) no adjustment is possible;

---- Torr is displayed.

NOTICE

Readjustment

Adjustment at atmospheric pressure

An adjustment at atmospheric pressure is only

possible if the pressure is higher than 525 Torr

(700 mbar).

Vent the measurement connection of the CVC 3000 E or in case that

of a connected optional external gauge head VSK 3000. Make sure

that the pressure transducer is at atmospheric pressure.

In ”Configuration” menu, select program ”Adjustment” at the controller.

Use the selection knob to adjust the reading to the current atmospheric

pressure.

Press the selection knob to confirm.

760 Torr

Note: To determine the actual atmospheric pressure, use an accurate

barometer or get accurate reading from the weather service, or a nearby

airport or other reliable source (taking into account the difference in alti-

tude between the source and the laboratory).

Advertising