Adjustment at atmospheric pressure

An adjustment at atmospheric pressure is only

possible in the upper pressure range.

➨

Vent the measurement connection of the VSP 3000 pressure trans-

ducer. Make sure that the measurement connection of the VSP 3000

is at atmospheric pressure. Allow the pressure transducer a period of

minimum 20 minutes to preheat at this pressure.

➨

In ”Configuration” menu, select program ”Adjustment” at the controller.

➨

The reading is automatically adjusted to ”750 Torr” (”1000 mbar”). This

value is fix and can not be changed.

➨

Press the selection knob to confirm.

750 Torr

➨

Allow the pressure transducer a period of minimum 20

minutes to preheat. Adjustment is not possible during

the preheat time (20 minutes after connecting the volt-

age supply).

➨

---- Torr is displayed, if adjustment is not possible in

the actual pressure range.

Adjustment under vacuum

An adjustment under vacuum is only possible

in the lower pressure range.

➨

Evacuate the measurement connection of the VSP 3000 pressure

transducer preferably below 7.5x10

-4

Torr (1x10

-3

mbar), e.g., by ap-

plying a high vacuum pump. Allow the pressure transducer a period of

minimum 20 minutes to preheat at this pressure.

➨

In ”Configuration” menu, select program ”Adjustment” at the controller.

+

The reading is automatically adjusted to ”zero”. This value is fix and

can not be changed.

➨

Press the selection knob to confirm.

0 Torr