Notes on operation

(1)

Connect the cable to the vacuum gauge head‘s serial in-

terface and to one the VACUU•BUS sockets (1) at the rear

of the vacuum gauge DCP 3000.

Attention: Do not cant when assembling or removing

plug connections! Comply with correct positioning of

the plug.

Only connect/disconnect the cable to/from the vacuum

gauge head when it is off circuit. Separate the DCP 3000

from mains.

Switch the DCP 3000 on.

When switching on the DCP 3000, a connected vacuum gauge head MPT 100 is automatically detected.

The vacuum gauge head is displayed as active sensor in the status line of the vacuum gauge as MPT

100. Over the whole measuring range the pressure reading is indicated in exponential notation.

Once all cables are connected, the vacuum gauge head MPT 100 is ready to operate. Allow the MPT

100 a stabilization time of approx. 5-10 minutes to meet its specifications. Even in case of the vacuum

gauge being switched off, the MPT 100 stays ready for operation. To de-energize the pressure vacuum

gauge head, unplug either its VACUU•BUS line or the power supply of the vacuum gauge.

The interior of the vacuum gauge head is highly sensitive! Do not insert fingers or

tools into the measuring chamber.

The measured pressure is dependent on the gas type (see ”Notes on operation”).

Max. permitted pressure at the pressure transducer: 1.5 bar (absolute).

Attention: Maximum pressure reading: 1*10

3

mbar. Pressure values above 1000

mbar are not displayed.

Risk of unnoticed overpressure. Risk of bursting!

NOTICE

Connecting several pressure transducers

At most, one vacuum gauge head MPT 100 can be connected to the vacuum gauge DCP 3000 from

software version 2.03 and up to four MPT 100 from DCP 3000 software version 2.22. However, it is pos-

sible to connect additionally several external pressure transducers VSK 3000 or VSP 3000 simultane-

ously to the DCP 3000.

Only the readings of one pressure transducer are displayed at a time.

The VACUU•BUS address of the active and displayed pressure transducer is shown in the upper left

corner (status line). The readings of all pressure transducers are shown in the menu ”Sensors” (see

instructions for use of DCP 3000).

Each pressure transducer needs an individual address (VSK 1 to VSK 4 / MPT 1 to MPT 4 (replacing

VSP 3000 if connected) / VSP 1 to VSP 4) to communicate with the DCP 3000 via VACUU•BUS.

Factory set each pressure transducer is set to address “1”.

To use several pressure transducers it is necessary to configurate the transducer appropriately.