Cleaning the mpt 100, Important notes concerning bake-out, Notice – VACUUBRAND MPT 200 User Manual

Page 14

Advertising
background image

page 14 of 20

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999226 // 07/10/2011

Cleaning the MPT 100

NOTICE

The vacuum gauge DCP 3000 itself is maintenance free.

In case of contamination inside the vacuum chamber (oil, particles, etc.) the vacuum gauge head may

be contaminated as well. Contamination of the vacuum gauge head MPT 100 or deposits will influence

the accuracy of measurement and/or the adjustment.

Attention: Never use a spiky or sharp-edged tool to clean the vacuum gauge

head!

The interior of the vacuum gauge head is highly sensitive! Do not insert fingers or

tools into the measuring chamber.

Clean a contaminated vacuum gauge head as follows:

Ensure that the vacuum gauge head is off circuit, disconnect the VACUU•BUS cable. Dismount the

vacuum gauge head from the vacuum chamber.

Fill the measurement chamber with a neutral solvent (e. g. benzene) and allow sufficient cleaning

time. Observe all regulations concerning usage and disposal of solvents!

Drain the solvent and dispose of in accordance with regulations, repeat cleaning if necessary.

Rinse the measurement chamber several times with alcohol in order to remove all solvent residues.

Allow the vacuum gauge head to dry, readjust if necessary.

Important notes concerning bake-out

The electronics and the magnetic unit of the vacuum gauge head MPT 100 can be

damaged during bake-out!

For baking out temperatures > 60°C, always remove the electronics and the mag-

netic unit from the sensor head. Bake out only the sensor head.

Pay attention to the components‘ assignments to each other, do not inter-

change.

Persons with active body aids (e.g. cardiac pacemakers) must keep a minimum

safety distance of approx. 10cm (4in) to the magnetic unit of the MPT 100 (see

also safety symbols on the vacuum gauge head).

Dismounting the electronics and the magnetic unit:

Loosen both set screws (fixating the sensor head to the housing of the electron-

ics) at the housing edges with a hex key size 2.

Pull the housing with the electronics from the sensor head.

Remove the magnetic unit from the sensor head.

Bake out the sensor head.

Mounting the electronics and the magnetic unit:

Put the magnetic unit onto the sensor head.

Mind the guidance given by the given groove.

Mount the electronics at the sensor head.

Mind the guidance given by the given groove.

Tighten the set screws (hex key size 2). Pay attention to a concise position be-

tween sensor head and magnetic unit.