Readjustment of the transmitter mpt 100, Readjustment of the vacuum gauge head mpt 100, Notice – VACUUBRAND MPT 200 User Manual

Page 16

Advertising
background image

page 16 of 20

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999226 // 07/10/2011

The vacuum gauge head MPT 100 has been adjusted in the works. Depending on

the process and/or on accuracy requirements, check the adjustment and readjust

if necessary (e.g. in case of contamination, different orientation, extreme tempera-

tures, etc.). For readjustment, the device has to be adjusted both at atmospheric

pressure as well as under vacuum.

For readjustment, install the vacuum gauge head in the same orientation in which

it will be operated later on at the application.

Allow the vacuum gauge head a period of minimum 5-10 minutes to preheat.

If no adjustment is possible in the current pressure range, ”---- mbar” is displayed.

Adjustment at atmospheric pressure

An adjustment at atmospheric pressure is only possible in the

upper pressure range.

Ventilate the vacuum gauge head MPT 100 and/or the vacu-

um system with air or N

2

. Make sure that the sensor head at

the MPT 100 is at atmospheric pressure. Allow the vacuum

gauge head a period of minimum 10 minutes to stabilize at

this pressure.

In function ”Configuration” select the menu item ”Adjust-

ment” (see instructions for use of DCP 3000).

The reading is automatically adjusted to ”1000 mbar”. This

value is fix and can not be changed.

Press the selection knob to confirm.

Adjustment under vacuum

Manual adjustment:

Evacuate the vacuum gauge head MPT 100 to a pressure be-

low 1*10

-5

mbar (e. g. by applying a high vacuum pump). Allow

the vacuum gauge head a period of minimum 10 minutes to

preheat at this pressure.

In function ”Configuration” select the menu item ”Adjust-

ment” (see instructions for use of DCP 3000).

The reading is automatically adjusted to ”zero”. This value

is fix and can not be changed.

Press the selection knob to confirm.

1000 mbar

0 mbar

Readjustment of the vacuum gauge head MPT 100

NOTICE

An

automatic adjustment of the Pirani sensor is run automatically, when the following conditions apply

simultaneously:

- the pressure measured by the cold cathode is < 1*10

-5

mbar.

- the measured value of the Pirani sensor does not change for one minute and does deviate too much

from the most recent calibration value.