Maintenance guide – Winco Mfg. S300 - Stretchair User Manual

Page 15

Advertising
background image

MAINTENANCE GUIDE

Monarch Hydro Power Unit Replacement

1. Place chair in the layflat position

(SEE PAGE.7)

with the vertical lift in its LOWEST position.

(FIG.15)

2. Unplug the Monarch Power cord from the battery

control box.

(FIG.16)

3. Locate the Power Unit positioned under the

upholstered seat.

(FIG. 17)

4. Remove the 5 screws and ALL nylon ties that `

secure the Power Unit to the chair frame.

5. Lift, rotate, and remove unit from the chair frame.

Be sure to keep the Power Unit vertical (brass

fittings up) to avoid oil spilliage.

6. Carefully unscrew all 4 brass fittings from the

Power Unit.

(FIG. 19)

7. Dispose of the old Power Unit carefully.

8. Cut black tubing to remove old ferrule and threaded

fitting. (Cut where shown in

FIG. 18)

9. Slide the black tube through the new threaded

fitting

10. Insert tubing into the open Hydro Power Unit port

11. Slide the new threaded fitting onto the Hydro Power

Unit port and tighten to a snug fit.

(FIG. 19)

12. Repeat steps 8-11 for all 4 port fittings

13. Re-mount Power Unit to base frame

14. Insert the Monarch Power cord into the battery

control box.

(FIG. 16)

15. Test the new unit by running the vertical lift up and

down 2-3 times.

FIG. 15

FIG. 16

FIG. 17

FIG. 18

FIG. 19

15