Guidance and manufacturer’s declaration –electromagnetic immunity

The Stretchair is intended for use in the electromagnetic environment specified below. The customer or the user of the Stretchair

should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment

guidance

Electrostatic discharge

(ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, concrete

or ceramic tile. If floors are

covered with synthetic material,

the relative humidity should be at

least 30 %.

Electrical fast

transient/burst

IEC 61000-4-4

±2 kV for power supply lines

±1 kV for input/output lines

±2 kV for power supply lines

±1 kV for input/output lines

Mains power quality should be

that of a typical commercial or

hospital environment.

Surge

IEC 61000-4-5

±1 kV differential mode

±2 kV common mode

±1 kV differential mode

±2 kV common mode

Mains power quality should be

that of a typical commercial or

hospital environment.

Voltage dips, short

interruptions and

voltage variations on

power supply input

lines IEC 61000-4-11

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 sec

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 sec

Mains power quality should be

that of a typical commercial or

hospital environment. If the user

of the Stretchair requires

continued operation during

power mains interruptions, it is

recommended that the Stretchair

be powered from an

uninterruptible power supply or a

battery.

Power frequency

(50/60 Hz) magnetic

field IEC 61000-4-8

3 A / m

3 A / m

Power frequency magnetic fields

should be at levels characteristic

of a typical location in a typical

commercial or hospital

environment.

NOTE U

T

is the a.c. mains voltage prior to application of the test level.