7 trigger export, 4 trigger modes, 1 post trigger mode – ADLINK PXIe-9852 User Manual

Page 31: 2 delayed trigger mode, Trigger export, Trigger modes, Post trigger mode, Delayed trigger mode, Figure 3-5, Post-trigger acquisition

Advertising
background image

Operations

21

PXIe-9852

3.3.7

Trigger Export

When acquisition is initiated, a pulse synchronized with the Time-
base clock asserts and is output through trigger output, at a pulse
width programmable from 50ns to 10μs via software.

3.4 Trigger Modes

Trigger modes applied to trigger sources initiate different data

acquisition timings when a trigger event occurs. The following trig-
ger mode descriptions are applied to analog input function.

3.4.1

Post Trigger Mode

Post-trigger acquisition is applicable when data is to be collected
after the trigger event, as shown. When the operation starts, PXIe-
9852 waits for a trigger event. Once the trigger signal is received,
acquisition begins. Data is generated from ADC and transferred to
system memory continuously. The acquisition stops once the total
data amount reaches a predefined value.

Figure 3-5: Post-Trigger Acquisition

3.4.2

Delayed Trigger Mode

Delayed-trigger acquisition is utilized to postpone data collection
after the trigger event, as shown. When PXIe-9852 receives a trig-
ger event, a time delay is implemented before commencing acqui-
sition. The delay is specified by a 16-bit counter value such that a

Advertising