Figure 3-13, 28 operations – ADLINK PXIe-9852 User Manual
Page 38

28
Operations
in width, therefore the lowest sampling rate is 3.051KS/s (200MS/s
/ 65535).
Figure 3-13: Varying Sampling Rates by Adjusting Scan Interval Counter
Counter
Name
Length
Valid Value
Description
ScanIntrv
16-bit
1-65535
Timebase divider to achieve
equivalent sampling rate of the
digitizer, where Sampling rate =
Timebase / ScanIntrv
DataCnt
28-bit
1-268435452
Specifies the amount of data to
be acquired:
X
1 - 268435452 for pre-
trig or mid-trig mode
operation
X
1 - 268435452 for Data
Average mode for 1
channel
X
1 - 134217724 for Data
Average mode for 2
channel
trigDelayTicks 16-bit
1 -65535
Indicates time between a trigger
event and commencement of
acquisition. The unit of a delay
count is the period of the
Timebase.
TIMEBASE
DATA
D1
Acquisition
In Progress
Trigger
Acquisition is initiated following this clock edge
D1
ScanIntrv = 1
D2
D3
D4
D5
D6
D7
D8
D9
D10
D2
D3
D4
D5
D6
D1
D2
D3
D4
ScanIntrv = 2
ScanIntrv = 3