2 applications, 3 specifications, 1 analog input – ADLINK PXIe-9529 User Manual
Page 12: Applications, Specifications, Analog input
2
Introduction
1.2 Applications
X
Structural health monitoring
X
Phase array data acquisition
X
Noise, vibration, and harshness (NVH) detection
X
Machine status monitoring
1.3 Specifications
1.3.1
Analog Input
Channel Characteristics
Comment
Channels
8
Type
Differential or Pseudo-Differential
Coupling
AC or DC, software selectable
AC coupling cutoff
frequency
0.5Hz
ADC resolution
24-Bit
ADC type
Delta-sigma
Input signal range ±10V, ±1V
Sampling rate (fs)
8 kS/s to 192 kS/s,
768 μS/s increments for fs > 108 kS/s,
576 μS/s increments for 54 kS/s ≤ fs
≤108 kS/s
Over voltage
protection
Differential: ±42.4V,
Pseudo-differential:
X
positive terminal ±42.4 V
X
negative terminal unpro-
tected, rated at ±2.5 V
Input impedance
1MΩ, (50Ω between negative input
and system ground for
pseudo-differential mode)
Offset error
±1 mV max.
Gain error
±0.1% of FSR