Section 3 – Installation

Dukane Manual Part No. 403-583-00

Power Cords

200/240 Volt Systems

The IEC AC power inlet connector mounted on the rear
panel requires a properly configured IEC compliant power
cord.

The 200/240 AC power cords supplied with the
generators are matched to the ultrasonic output power
rating and the continent of specified use. See Table 3-I.

Table 3- I

Standard IEC AC Power Cord Part Numbers

100/120 Volt Systems

(North America or Japan)

The power cord (including strain relief) supplied with the
100/120 AC systems is permanently attached to the rear
of the generator. Units with this power cord are for use in
North America or Japan.

Automation Controlled System

Step 1. Ground the generator chassis using the supplied

14-Gauge wire, and attach it to the grounding
stud:

C

in Figure 3 -3.

Step 2. Optional – Ground the probe support. This is a

user–supplied 14-Gauge wire.

Step 3. Input/Output Cable - Attach the automation

control cable from the user–supplied automa-
tion equipment to the system HD-15 connector,
INPUTS/OUTPUTS on the rear panel:

B

in

Figure 3 -3.

Step 4. Attach the high voltage coaxial cable from the

probe to the ultrasound output connector,

D

in

Figure 3 -3.

Step 5. Connect the AC power cord to the generator

IEC power inlet connector, and plug the other
end into an approved AC outlet:

A

in Figure 3 -3.

Continent of Use

Power Cord

Part Number

Power

North America

200 - 1541

240V, 10A

Europe

200 - 1542

240V, 10A

India

200 - 1624

240V, 10A