Section 7 – Troubleshooting

Page  5

Dukane Manual Part No. 403-583-00

No Ultrasonic Output

Probe

Make sure that the probe cable is connected to the generator
connector (PROBE) and secured to the rear panel. Also,
make sure the probe stack is properly assembled.

System Power Output Level

Overload

When an overload occurs, it will automatically reset
when the next ultrasound activation signal begins. If the
condition persists:

Turn the generator OFF and:
1. Check the system. Change the probe to one that is

known to be good.

2. Turn the generator

ON, and see if the fault condition

has been corrected.

Overtemperature

When the system overheats, and the generator’s internal
temperature exceeds 85°C (185° F) an overtemperature
fault condition will trip.
When the system cools, the system automatically resets
the overtemperature fault.

Generator Fault Does Not Reset

When a system fault condition does not automatically
reset, the generator needs servicing.

System Power Diagnostic
Procedures

The only fault indications available with these probe
systems are the ones shown on the LCD display.

NOTE

The LCD screen displays a variety of
pop-up status changes as they oc-
cur. Check Table 7-I - Pop-up Fault
Status Screens - Page 53.

NOTE

When Latching Faults is enabled,
ENTER must be pressed to clear
a fault.