Hanna Instruments HI 98186 User Manual

Page 22

Advertising
background image

22

If the DO reading is less than the minimum end DO value set during OUR
configuration a warning icon will be displayed and a beep will be heard
every two seconds. Press Stop to stop the test and the beeper.

To end the OUR test before the maximum time interval set during OUR
configuration press Stop.
If Stop is pressed before the minimum time for the OUR test has elapsed, the
instrument will display a warning message.

Press Resume to continue the test or Stop to end the test.
At the end of the test the meter will display the computed OUR value and the
duration of the test.
Press LOG to save a complete set of data regarding the OUR test.

Press Start to begin a new OUR test.

Notes:

If at the end of the test the DO reading is less than the minimun end DO

value set during OUR configuration, a warning message will be dis-
played.

Press any key to clear the message from the screen, or press HELP to view
detailed information about the warning.

Advertising