26
At the end of the test the meter will display the computed SOUR value and
the duration of the test.
Press LOG to save a complete set of data regarding the SOUR test.
Press Start to begin a new SOUR test.
Notes:
•
If the DO reading is less than the minimun end DO value set during SOUR
configuration, a warning message will be displayed.
Press any key to clear the message from the screen, or press HELP to view
detailed information about the warning.
•
If the DO value is greater than the DO value from the beginning of the test
an error message will be displayed.
Press Start to begin a new SOUR test or ESC to return to the SOUR
measure screen.