Hanna Instruments HI 98186 User Manual

Page 55

Advertising
background image

55

CHR xx

Change the instrument range according with the parameter
value (xx):
• xx=20 DO range

• xx=21 BOD range

• xx=22 OUR range
• xx=23 SOUR range
The instrument will answer for these commands with:

<STX> <answer

> <ETX>

where:

<STX> is 02 ASCII code character (start of text)
<ETX> is 03 ASCII code character (end of text)
<answer>:
<ACK> is 06 ASCII code character (recognized command)
<NAK> is 21 ASCII code character (unrecognized command)
<CAN> is 24 ASCII code character (corrupted command)

COMMANDS REQUIRING AN ANSWER
The instrument will answer for these commands with:

<STX> <answer> <checksum> <ETX>

where the checksum is the bytes sum of the answer string sent as 2 ASCII characters.
All the answer messages are with ASCII characters.
RAS Causes the instrument to send a complete set of readings in according

with the current range:

DO, temperature, and pressure on the DO and BOD range

DO, temperature, pressure, OUR/SOUR value, OUR/SOUR test time

on the OUR/SOUR range

BOD result, start DO value and end DO value when in the BO result

screen (meter mode 25)

The answer string contains:

Meter mode (2 chars):

- 20 - DO range
- 21 - BOD range
- 22 - OUR range
- 23 - SOUR range
- 25 - BOD result screen

Meter status (2 chars of status byte): represents a 8 bit hexadeci-

mal encoding