Hanna Instruments HI 98186 User Manual

Page 56

Advertising
background image

56

0x10: temperature probe connection

0x20: DO measure unit (0 = %, 1= mg/L)

0x01: new GLP

GLP

GLP

GLP

GLP data available

0x02: new setup parameter

0x04: out of calibration range

0x08: the meter is in autoend mode;

Reading status: R - in range, O - over range, U - under range

The following status flags are sent for all modes except

mode 25

DO reading range flag

temperature reading range flag

pressure reading range flag

OUR/SOUR reading range flag (sent only if in OUR/SOUR

measure mode)

The following status flag is sent only for mode 25

BOD reading range flag

Readings

The following values are sent for all modes except

mode 25

DO reading, including sign and decimal point (8 chars)

temperature, including sign and decimal point (8chars)

pressure value, including sign and decimal point, always

in mmHg (11 chars)

OUR/SOUR reading, including sign and decimal point

(sent only if in OUR/SOUR measure mode) (8 chars)

OUR/SOUR counter (sent only if in OUR/SOUR measure

mode) (4 chars)

The following values are sent only for mode 25

BOD reading, including sign and decimal point (8 chars)

initial DO value, including sign and decimal point [mg/L]

(6 chars)

final DO value, including sign and decimal point [mg/L]

(6 chars)

MDR Requests the instrument model name and firmware code (16 ASCII chars).
GLP Requests the calibration data record.
The answer string contains:

the number of calibrated buffers (1 char)

calibrated buffer unit (0 = %, 1 = mg/L) (1 char)

buffer value including sign and decimal point (6 chars)

calibrated buffer unit (0 = %, 1 = mg/L) (this value is sent only

if there is a 2 point calibration) (1 char)

Advertising