Hanna Instruments HI 98186 User Manual

Page 56

Advertising
background image

56

•

0x10: temperature probe connection

•

0x20: DO measure unit (0 = %, 1= mg/L)

•

0x01: new GLP

GLP

GLP

GLP

GLP data available

•

0x02: new setup parameter

•

0x04: out of calibration range

•

0x08: the meter is in autoend mode;

•

Reading status: R - in range, O - over range, U - under range

•

The following status flags are sent for all modes except

mode 25

•

DO reading range flag

•

temperature reading range flag

•

pressure reading range flag

•

OUR/SOUR reading range flag (sent only if in OUR/SOUR

measure mode)

•

The following status flag is sent only for mode 25

•

BOD reading range flag

•

Readings

•

The following values are sent for all modes except

mode 25

•

DO reading, including sign and decimal point (8 chars)

•

temperature, including sign and decimal point (8chars)

•

pressure value, including sign and decimal point, always

in mmHg (11 chars)

•

OUR/SOUR reading, including sign and decimal point

(sent only if in OUR/SOUR measure mode) (8 chars)

•

OUR/SOUR counter (sent only if in OUR/SOUR measure

mode) (4 chars)

•

The following values are sent only for mode 25

•

BOD reading, including sign and decimal point (8 chars)

•

initial DO value, including sign and decimal point [mg/L]

(6 chars)

•

final DO value, including sign and decimal point [mg/L]

(6 chars)

MDR Requests the instrument model name and firmware code (16 ASCII chars).
GLP Requests the calibration data record.
The answer string contains:

•

the number of calibrated buffers (1 char)

•

calibrated buffer unit (0 = %, 1 = mg/L) (1 char)

•

buffer value including sign and decimal point (6 chars)

•

calibrated buffer unit (0 = %, 1 = mg/L) (this value is sent only

if there is a 2 point calibration) (1 char)