Use the ARROW keys to set the calibration point values to that of the ice

and water mixture, measured by the reference thermometer.

When the reading is stable and close to the selected calibration point, the

CFM functional key is displayed.

Press CFM to confirm.

• The second expected calibration point is displayed.

Immerse the DO probe into the second vessel as close as possible to the

reference thermometer.

Allow a few seconds for the probe to stabilize.

Use the ARROW key to set the calibration point value to that measured

by the reference thermometer.