Use the ARROW keys to set the calibration point values to that of the ice
and water mixture, measured by the reference thermometer.
•
When the reading is stable and close to the selected calibration point, the
CFM functional key is displayed.
•
Press CFM to confirm.
• The second expected calibration point is displayed.
•
Immerse the DO probe into the second vessel as close as possible to the
reference thermometer.
•
Allow a few seconds for the probe to stabilize.
•
Use the ARROW key to set the calibration point value to that measured
by the reference thermometer.