Setup – Hanna Instruments HI 98186 User Manual

Page 33

Advertising
background image

33

Setup mode allows viewing and modifying the measurement parameters.
The following table lists the SETUP parameters, their valid range and the

factory default settings.

Item

Description

Valid Value

Default

Calibr. Time-out Number of days after

Disabled, 1 to 7 days

Disabled

calibr. warning is displayed

Salinity

The solution’s salt content 0 to 70 g/L

0 g/L

BOD
Configuration
-Sample min The minimum diff. between 0.00 to 50.00 mg/L

0.00 mg/L

delta DO

the start and the end DO val.

-Sample min The minimum end DO value 0.00 to 50.00 mg/L

0.00 mg/L

end DO
-Seed min The minimum diff. between 0.00 to 50.00 mg/L

0.00 mg/L

delta DO

the start and the end DO val.

-Seed min The minimum end DO value 0.00 to 50.00 mg/L

0.00 mg/L

end DO
OUR
Configuration
-Min time

The minimum time for the 1 to 3600 sec.

1 s

OUR test

-Max time

The maximum time for the 1 to 3600 sec.

3600 sec.

OUR test

-Min start DO The minimum DO value 0.01 to 50.00 mg/L

0.01 mg/L

for starting the OUR test

-Min end DO The minimum DO value at 0.00 to 50.00 mg/L

0.00 mg/L

the end of the OUR test

-Total volume The total volume of the 0.1 to 300.0 mL

0.1 mL

solution to be tested

-Sample volume The volume of sample in 0.1 to 300.0 mL

0.1 mL

the solution to be tested

SOUR
configuration
-Min time

The minimum

1 to 3600 sec.

1 sec.

time for the SOUR test

-Max time

The maximum
time for the SOUR test

1 to 3600 sec.

3600 sec.

-Min start DO The minimum DO value

for starting the SOUR test 0.01 to 50.00 mg/L

0.01 mg/L

-Min end DO The minimum DO value at 0.00 to 50.00 mg/L

0.00 mg/L

the end of the test

SETUP

SETUP

SETUP

SETUP

SETUP