Hanna Instruments HI 98186 User Manual

Page 57

Advertising
background image

57

buffer value including sign and decimal point (this value is sent only

if there is a 2 point calibration) (6 chars)

salinity value (3 chars)

pressure value in mmHg, including sign and decimal point

(11 chars)

temperature value including sign and decimal point (8 chars)

calibration time: yymmddhhmmss (12 chars).

PAR Requests the setup parameters setting.

The answer string contains:

backlight value (1 char)

contrast value (2 chars)

instrument ID (4 chars)

calibration alarm timeout (2 chars)

Setup information (2 chars): 8 bit hexadecimal encoding:

0x01: -

-

-

-

- beep is on

0x04: -

-

-

-

- Celsius / Fahrenheit display (°C if the bit is set)

0x10: - manual pressure (1 if activated, 0 otherwise)

Auto Light-off time (3 chars)

Auto Power-off time (3 chars)

Salinity value (3 chars)

Pressure unit (1 char): 0 - mmHg, 1 - inHg, 2 - atm, 3 - mbar,

4 - psi, 5 - kPa

BOD configuration values

sample minimum delta DO, including sign and decimal point

(6 chars)

sample minimum end DO, including sign and decimal point

(6 chars)

seed minimum delta DO, including sign and decimal point

(6 chars)

seed minimum end DO, including sign and decimal point

(6 chars)

OUR configuration values

minimum time in seconds (4 chars)

maximum time in seconds (4 chars)

minimum start DO including sign and decimal point (6 chars)

minimum end DO including sign and decimal point (6 chars)

total volume including sign and decimal point (6 chars)

sample volume including sign and decimal point (6 chars)

SOUR configuration values

minimum time in seconds (4 chars)