Hanna Instruments HI 98186 User Manual

Page 36

Advertising
background image

36

BOD configuration
Focus on

BOD configuration item.

Press Select.

Parameters:

Sample min

D

DO -the minimum acceptable difference between the

initial and final DO values for a sample. If the difference is less than this
value the meter will show a warning message when evaluating the BOD.

Range: 0.00 to 50.00 mg/L.

Sample min end DO - the minimum acceptable final DO value for a

sample. If the final DO value is less than this value the meter will show a
warning message when evaluating the BOD.

Range: 0.00 to 50.00 mg/L.

Seed min

D

DO - the minimum acceptable difference between the initial

and final DO values for a seed sample. If the difference is less than this
value the meter will show a warning message when evaluating the BOD.

Range: 0.00 to 50.00 mg/L.

Seed min end DO - the minimum acceptable final DO value is less than

this value the meter will show a warning message when evaluating the
BOD.

Range: 0.00 to 50.00 mg/L.
Press Prev/Next to select a different parameter.
Press ARROW keys to modify the selected parameter’s value.
Press Save to save the new BOD configuration.
Press ESC to leave without changing.