Chapter 3: Getting Started
Product Service 1-800-522-7658
12
Connecting to the Exhaust System (Optional on
XPert Filtered Balance System)
ATTENTION: The weight of the exhaust ductwork system must
be supported independently of the enclosure superstructure or
damage may occur.
The exhaust system should be installed by a qualified HVAC
contractor.
The exhaust connection on the 2', 3' and 4' XPert Filtered Balance Stations have
been designed to accept 6" diameter ductwork as shown in Figure 3-2. The 5' and
6' XPert Filtered Balance Stations have been designed for 10" diameter ductwork.
For the 2', 3' and 4' sizes, the exhaust connection ships uninstalled on the top and
needs to be fastened with the screws provided. See Chapter 7 for ordering Canopy
Exhaust Connection for the XPert Filtered Balance System. Both 6" for house
exhaust and 5" for connection to a FilterMate Portable Exhauster are available.
Canopy Exhaust Connection aids in the removal of chemicals or applications
where a higher degree of powder and particulate removal is required. Review
Chapter 2 for exhaust prerequisites and review Chapter 7 for ordering blower
exhaust equipment. See Figure 3-2 for the XPert Station and Figure 3-3 for
Canopy Exhaust Connection options for the XPert Filtered Balance System.
Consult Labconco Customer Service should you require help sizing your blower
for the exhaust volume and system static pressure loss.
To ensure compatibility, the selected exhaust duct material
should match the enclosure, procedures and chemical
applications.
!
!