Programming the guardian 1000 digital alarm – Labconco XPert Filtered Balance Stations 3962622 User Manual

Page 47

Advertising
background image

Chapter 6: Maintaining Your Filtered Enclosure

Product Service 1-800-522-7658

43


Operation

The Guardian Digital Airflow Monitor consists of the airflow sensor, the Alarm
Unit, and the 15 VDC power supply. For 115V operation the alarm unit is
powered by plugging the power supply into the factory-prepared digital airflow
monitor socket. For 230V operation, the Alarm Unit is powered by plugging the
power supply into a building outlet. The alarm has “Enter,” “+,” and “-” buttons to
program the monitor. There is also a green LED “SAFE,” yellow LED
“CAUTION,” and red LED “LOW” with audible alarm for airflow conditions.
The audible alarm can be permanently muted if desired. The Guardian Digital
1000 Airflow Monitor displays a face velocity value, provides an RS232
communications port to a PC or building computer system, can be configured for
external input connections such as night setback or external alarm and provides up
to three output relays that can be configured. For complete detailed information,
please refer to the separate Labconco 1000 Alarm User’s Manual provided with
the enclosure.

Calibration

The filtered enclosure comes factory calibrated with the monitor alarm at 60 ± 10
fpm with the inflow velocity set at 90± 10 fpm.

Programming the Guardian 1000 Digital Alarm

Set the desired face velocity of the enclosure with the sash in the down position
using an average face velocity of readings from a thermal anemometer.

a. 2 ft enclosure – 2 readings
b. 3 ft enclosure – 3 readings
c. 4 ft enclosure – 4 readings

Configure Alarm

To successfully calibrate the airflow monitor, it will be necessary to change the
face velocity by adjusting the airflow exhaust volume. The exhaust volume can be
adjusted using the speed control located behind the front panel on the XPert
Filtered System. For the XPert Filtered Station, the exhaust volume is adjusted
through a damper in the building exhaust system.

The following inflow face velocity speeds are recommended to successfully
calibrate. To begin, determine what the low airflow, low warning airflow, normal
airflow, and high airflow shall be.

Advertising
This manual is related to the following products:

XPert Filtered Balance Stations 3962621, XPert Filtered Balance Stations 3962620, XPert Filtered Balance Stations 3962602, XPert Filtered Balance Stations 3962600, XPert Filtered Balance Stations 3962522, XPert Filtered Balance Stations 3962521, XPert Filtered Balance Stations 3962520, XPert Filtered Balance Stations 3962502, XPert Filtered Balance Stations 3962501, XPert Filtered Balance Stations 3962500, XPert Filtered Balance Stations 3940422, XPert Filtered Balance Stations 3940421, XPert Filtered Balance Stations 3940420, XPert Filtered Balance Stations 3940402, XPert Filtered Balance Stations 3940401, XPert Filtered Balance Stations 3940400, XPert Filtered Balance Stations 3940321, XPert Filtered Balance Stations 3940320, XPert Filtered Balance Stations 3940302, XPert Filtered Balance Stations 3940301, XPert Filtered Balance Stations 3940300, XPert Filtered Balance Stations 3940222, XPert Filtered Balance Stations 3940221, XPert Filtered Balance Stations 3940220, XPert Filtered Balance Stations 3940202, XPert Filtered Balance Stations 3940201, XPert Filtered Balance Stations 3940200, XPert Filtered Balance Systems 3971402, XPert Filtered Balance Systems 3971401, XPert Filtered Balance Systems 3971400, XPert Filtered Balance Systems 3971302, XPert Filtered Balance Systems 3971301, XPert Filtered Balance Systems 3971300, XPert Filtered Balance Systems 3971202, XPert Filtered Balance Systems 3971201, XPert Filtered Balance Systems 3971200, XPert Filtered Balance Systems 3961622, XPert Filtered Balance Systems 3961621, XPert Filtered Balance Systems 3961620, XPert Filtered Balance Systems 3961602, XPert Filtered Balance Systems 3961601, XPert Filtered Balance Systems 3961600, XPert Filtered Balance Systems 3961522, XPert Filtered Balance Systems 3961521, XPert Filtered Balance Systems 3961520, XPert Filtered Balance Systems 3961502, XPert Filtered Balance Systems 3961501, XPert Filtered Balance Systems 3961500, XPert Filtered Balance Systems 3950422, XPert Filtered Balance Systems 3950421, XPert Filtered Balance Systems 3950420, XPert Filtered Balance Systems 3950402, XPert Filtered Balance Systems 3950401, XPert Filtered Balance Systems 3950400, XPert Filtered Balance Systems 3950322, XPert Filtered Balance Systems 3950321, XPert Filtered Balance Systems 3950320, XPert Filtered Balance Systems 3950302, XPert Filtered Balance Systems 3950301, XPert Filtered Balance Systems 3950300, XPert Filtered Balance Systems 3950222, XPert Filtered Balance Systems 3950221, XPert Filtered Balance Systems 3950220, XPert Filtered Balance Systems 3950202, XPert Filtered Balance Systems 3950201, XPert Filtered Balance Systems 3950200