Chapter 3: Getting Started
Product Service 1-800-522-7658
17
See Figure 3-4. The grounding connection shall not be made to the terminal box
cover. The building electrical supply system for the remote blower should include
overload protection such as a switch or circuit breaker in close proximity and
within easy reach of the operator. The switch or circuit breaker shall be marked as
the disconnecting device. Consult NEC-2002 for proper installation.
Figure 3-4
XPert Filtered Balance Station Conduit Location for Wiring External
Remote Blower.
2', 3', & 4' XPERT FILTRED
BALANCE STATION 6.06" ID
EXHAUST OUTLET TO REMOTE
BLOWER, 10.09” ID FOR 5' AND 6'
ROUTE CONDUIT THROUGH HOLE
PROVIDED IN FRONT PANEL