Chapter 6: Maintaining Your Filtered Enclosure

Product Service 1-800-522-7658

38

HEPA Filter Leak Test

Purpose

After installing the new HEPA filter, the HEPA filter should be leak checked.
This test is performed to determine the integrity of the HEPA filter, the filter
housing, and the filter mounting frames. Leak testing is to be done by a qualified
technician with calibrated equipment. Remove the top perforated exhaust cover by
using a Phillips screwdriver to unfasten the (2) screws used to secure (2) clips. See
Figure 6-1. The filter passes the leak test at .01% or better. Reference Leak
Testing and Photometer scanning from the Institute of Environmental Services
(IES-RP-CC001.3)

Apparatus

1. An aerosol photometer ATI model 2D, 2E, 2G or equivalent. Air

Techniques Hamilton Associates inc. 11403 Cron Ridge Dr. Owings Mills,
MD 21117

2. One aerosol generator of the Laskin nozzle(s) type. An aerosol of mineral

oil or suitable liquid shall be created by flowing air through it. The compressed
air supplied to the generator should be adjusted to a pressure of 10± 1 psig.
during operation. Air Techniques Inc. Model TDA-4A or equal. One nozzle at
10 psig is (67.5 cfm x 100ug/l)/(Vol. of air), For the 2' at 90 fpm or 130 cfm,
one nozzle @10 psig is 6,750/130 cfm = 52 ug/l. For the 3' at 90 fpm or 200
cfm, one nozzle @ 10 psig is 34 ug/l. For the 4' at 90 fpm or 265 cfm, one
nozzle at 10 psig is 26 ug/l. For the 5' at 90 fpm or 215 cfm, one nozzle at 10
psig is 31 ug/l. For the 6' at 90 fpm or 260 cfm, one nozzle at 10 psig is 26
ug/l. Note 5' and 6' use two filters.

3. Mineral oil (Catalog #1491400).

4. Sampling Nozzle, Rectangular 1/2" x 3-1/4", Air Techniques, Inc.

Procedure

For the ATI 2G Photometer

1. Turn on the photometer and allow it to operate for a minimum of 5 minutes.

Leave the valve in the “CLEAR” setting.

2. Press the “ENTER” keypad. Press the “REF” keypad.

3. The display will display “P1” for approximately 1 second, and then display a

numerical value.

4. Using the “^” or “ν” keypads, respectively, increase or decrease the numerical

value until it equals 52 (2'), 34 (3'), 26 (4'), 31 (5'), 26 (6') for the enclosure at
90 fpm inflow velocity.

Advertising
This manual is related to the following products:

XPert Filtered Balance Stations 3962621, XPert Filtered Balance Stations 3962620, XPert Filtered Balance Stations 3962602, XPert Filtered Balance Stations 3962600, XPert Filtered Balance Stations 3962522, XPert Filtered Balance Stations 3962521, XPert Filtered Balance Stations 3962520, XPert Filtered Balance Stations 3962502, XPert Filtered Balance Stations 3962501, XPert Filtered Balance Stations 3962500, XPert Filtered Balance Stations 3940422, XPert Filtered Balance Stations 3940421, XPert Filtered Balance Stations 3940420, XPert Filtered Balance Stations 3940402, XPert Filtered Balance Stations 3940401, XPert Filtered Balance Stations 3940400, XPert Filtered Balance Stations 3940321, XPert Filtered Balance Stations 3940320, XPert Filtered Balance Stations 3940302, XPert Filtered Balance Stations 3940301, XPert Filtered Balance Stations 3940300, XPert Filtered Balance Stations 3940222, XPert Filtered Balance Stations 3940221, XPert Filtered Balance Stations 3940220, XPert Filtered Balance Stations 3940202, XPert Filtered Balance Stations 3940201, XPert Filtered Balance Stations 3940200, XPert Filtered Balance Systems 3971402, XPert Filtered Balance Systems 3971401, XPert Filtered Balance Systems 3971400, XPert Filtered Balance Systems 3971302, XPert Filtered Balance Systems 3971301, XPert Filtered Balance Systems 3971300, XPert Filtered Balance Systems 3971202, XPert Filtered Balance Systems 3971201, XPert Filtered Balance Systems 3971200, XPert Filtered Balance Systems 3961622, XPert Filtered Balance Systems 3961621, XPert Filtered Balance Systems 3961620, XPert Filtered Balance Systems 3961602, XPert Filtered Balance Systems 3961601, XPert Filtered Balance Systems 3961600, XPert Filtered Balance Systems 3961522, XPert Filtered Balance Systems 3961521, XPert Filtered Balance Systems 3961520, XPert Filtered Balance Systems 3961502, XPert Filtered Balance Systems 3961501, XPert Filtered Balance Systems 3961500, XPert Filtered Balance Systems 3950422, XPert Filtered Balance Systems 3950421, XPert Filtered Balance Systems 3950420, XPert Filtered Balance Systems 3950402, XPert Filtered Balance Systems 3950401, XPert Filtered Balance Systems 3950400, XPert Filtered Balance Systems 3950322, XPert Filtered Balance Systems 3950321, XPert Filtered Balance Systems 3950320, XPert Filtered Balance Systems 3950302, XPert Filtered Balance Systems 3950301, XPert Filtered Balance Systems 3950300, XPert Filtered Balance Systems 3950222, XPert Filtered Balance Systems 3950221, XPert Filtered Balance Systems 3950220, XPert Filtered Balance Systems 3950202, XPert Filtered Balance Systems 3950201, XPert Filtered Balance Systems 3950200