Chapter 6: Maintaining Your Filtered Enclosure

Product Service 1-800-522-7658

35

Routine Maintenance Schedule

Weekly

Wipe down the interior surfaces of the enclosure with a disinfectant or

cleaner, depending upon the usage of the unit and allow to dry.

Using a damp cloth, clean the exterior surfaces of the enclosure,

particularly the front and top to remove any accumulated dust.

Operate the exhaust system, noting the airflow velocity through the

enclosure using a source of visible smoke. Airflow monitors are
recommended for constant monitoring.

Monthly (or more often as required)

Determine the actual face velocity through the sash opening of the

enclosure where the average reading should be at the specified velocity.
(Use calibrated thermal anemometer or other approved apparatus). Airflow
alarms are recommended for constant monitoring.

The enclosure rear baffle should be checked for any blockage to ensure that

the enclosure is maintaining proper airflow.

All weekly activities.

Check face velocity. Increase speed control or change HEPA filter when

face velocity of the enclosure drops below the recommended speed for your
facility or if the airflow alarm monitor alerts you. Airflow monitors are
recommended.

While the enclosure is filled with the contaminant, test filter condition on

Odor Control carbon filters using the appropriate gas detector tube at
intervals of 20% of the total estimated time. The exception to the 20%
recommendation is formaldehyde or any carcinogen or suspected
carcinogen. These hazardous chemicals must be checked at least every
10% of the total estimated time. Gas detector tubes for the specific
chemicals that are being used in the enclosure can be obtained from your
laboratory supply dealer.

Replace Odor Control carbon filters when chemical breakthrough is

indicated by odor, time, detector tube, or for some chemicals, analytical
instrumentation. See “Replacing Odor Control Carbon Filters” section of
this manual in Chapter 6.

Annually

Replace the fluorescent lamps.

Have the enclosure validated by a qualified certification technician. See

Certification and Recertification in Chapter 6.

All monthly activities.


HEPA

CARBON

CARBON

Advertising
This manual is related to the following products:

XPert Filtered Balance Stations 3962621, XPert Filtered Balance Stations 3962620, XPert Filtered Balance Stations 3962602, XPert Filtered Balance Stations 3962600, XPert Filtered Balance Stations 3962522, XPert Filtered Balance Stations 3962521, XPert Filtered Balance Stations 3962520, XPert Filtered Balance Stations 3962502, XPert Filtered Balance Stations 3962501, XPert Filtered Balance Stations 3962500, XPert Filtered Balance Stations 3940422, XPert Filtered Balance Stations 3940421, XPert Filtered Balance Stations 3940420, XPert Filtered Balance Stations 3940402, XPert Filtered Balance Stations 3940401, XPert Filtered Balance Stations 3940400, XPert Filtered Balance Stations 3940321, XPert Filtered Balance Stations 3940320, XPert Filtered Balance Stations 3940302, XPert Filtered Balance Stations 3940301, XPert Filtered Balance Stations 3940300, XPert Filtered Balance Stations 3940222, XPert Filtered Balance Stations 3940221, XPert Filtered Balance Stations 3940220, XPert Filtered Balance Stations 3940202, XPert Filtered Balance Stations 3940201, XPert Filtered Balance Stations 3940200, XPert Filtered Balance Systems 3971402, XPert Filtered Balance Systems 3971401, XPert Filtered Balance Systems 3971400, XPert Filtered Balance Systems 3971302, XPert Filtered Balance Systems 3971301, XPert Filtered Balance Systems 3971300, XPert Filtered Balance Systems 3971202, XPert Filtered Balance Systems 3971201, XPert Filtered Balance Systems 3971200, XPert Filtered Balance Systems 3961622, XPert Filtered Balance Systems 3961621, XPert Filtered Balance Systems 3961620, XPert Filtered Balance Systems 3961602, XPert Filtered Balance Systems 3961601, XPert Filtered Balance Systems 3961600, XPert Filtered Balance Systems 3961522, XPert Filtered Balance Systems 3961521, XPert Filtered Balance Systems 3961520, XPert Filtered Balance Systems 3961502, XPert Filtered Balance Systems 3961501, XPert Filtered Balance Systems 3961500, XPert Filtered Balance Systems 3950422, XPert Filtered Balance Systems 3950421, XPert Filtered Balance Systems 3950420, XPert Filtered Balance Systems 3950402, XPert Filtered Balance Systems 3950401, XPert Filtered Balance Systems 3950400, XPert Filtered Balance Systems 3950322, XPert Filtered Balance Systems 3950321, XPert Filtered Balance Systems 3950320, XPert Filtered Balance Systems 3950302, XPert Filtered Balance Systems 3950301, XPert Filtered Balance Systems 3950300, XPert Filtered Balance Systems 3950222, XPert Filtered Balance Systems 3950221, XPert Filtered Balance Systems 3950220, XPert Filtered Balance Systems 3950202, XPert Filtered Balance Systems 3950201, XPert Filtered Balance Systems 3950200