Chapter 6: Maintaining Your Filtered Enclosure

Product Service 1-800-522-7658

48

Time - For applications that have very consistent inlet concentrations and
operating time, “Time” can be used to anticipate saturation or TWA levels based
on prior experience. However, this does not replace the need for sampling.
Consult Labconco technical specialist for an estimate of carbon filter life based on
chemical usage. Detector tubes, or analytical instrumentation should always be
used to determine concentrations in the carbon filter. It is recommended that the
carbon filters be checked with detector tubes or other means at intervals of 20% of
the total estimated filter life. The exception to the 20% recommendation is
formaldehyde and any carcinogen or suspected carcinogen. These more hazardous
chemicals must be checked at least every 10% of the total estimated time.

Analytical Instrumentation - This is the most accurate means of measuring
concentrations of any chemical. It is the method of choice when no detector tubes
are available or the tubes are not sensitive enough to measure at the TWA
concentration for the chemical. This method is also to be used to determine
saturation when the chemical concentration is below the measurement range of
detector tubes.

Odor Control Carbon Filter Replacement Procedure – See Figures 4-1 and 4-2
for Carbon Filter Replacement.

1. The carbon filters are replaced by first removing the two screws, which

hold the clips to secure the upper diffuser screen. Remove the screws, clips
and upper diffuser screen.

2. Remove the carbon filter in a careful way to avoid the flow switch and

HEPA filter.

3. Re-install the new carbon filter with the gasket down. Replace the upper

diffuser screen, clips and two screws.

4. The weight of the carbon filter with the gasket down will compress the

gasket.

Calculating Odor Control Carbon Filter Life

Labconco developed a modeling program to estimate the filter life for typical
carbon filters. Since filter life is dependent on the chemical used, the airflow, filter
size, and the dwell time, refer to the Chemical Guide for the Paramount® Filtered
Enclosure. The estimated life for Odor Control carbon filters for the filtered
enclosures is conservatively calculated at 50% or half of the published values for
the Paramount in the Chemical Guide. For example, if you use isopropyl alcohol
to disinfect and use approximately 100 ml per week during 2 hours of use per day
then follow these steps to calculate the concentration in parts per million (ppm).

Advertising
This manual is related to the following products:

XPert Filtered Balance Stations 3962621, XPert Filtered Balance Stations 3962620, XPert Filtered Balance Stations 3962602, XPert Filtered Balance Stations 3962600, XPert Filtered Balance Stations 3962522, XPert Filtered Balance Stations 3962521, XPert Filtered Balance Stations 3962520, XPert Filtered Balance Stations 3962502, XPert Filtered Balance Stations 3962501, XPert Filtered Balance Stations 3962500, XPert Filtered Balance Stations 3940422, XPert Filtered Balance Stations 3940421, XPert Filtered Balance Stations 3940420, XPert Filtered Balance Stations 3940402, XPert Filtered Balance Stations 3940401, XPert Filtered Balance Stations 3940400, XPert Filtered Balance Stations 3940321, XPert Filtered Balance Stations 3940320, XPert Filtered Balance Stations 3940302, XPert Filtered Balance Stations 3940301, XPert Filtered Balance Stations 3940300, XPert Filtered Balance Stations 3940222, XPert Filtered Balance Stations 3940221, XPert Filtered Balance Stations 3940220, XPert Filtered Balance Stations 3940202, XPert Filtered Balance Stations 3940201, XPert Filtered Balance Stations 3940200, XPert Filtered Balance Systems 3971402, XPert Filtered Balance Systems 3971401, XPert Filtered Balance Systems 3971400, XPert Filtered Balance Systems 3971302, XPert Filtered Balance Systems 3971301, XPert Filtered Balance Systems 3971300, XPert Filtered Balance Systems 3971202, XPert Filtered Balance Systems 3971201, XPert Filtered Balance Systems 3971200, XPert Filtered Balance Systems 3961622, XPert Filtered Balance Systems 3961621, XPert Filtered Balance Systems 3961620, XPert Filtered Balance Systems 3961602, XPert Filtered Balance Systems 3961601, XPert Filtered Balance Systems 3961600, XPert Filtered Balance Systems 3961522, XPert Filtered Balance Systems 3961521, XPert Filtered Balance Systems 3961520, XPert Filtered Balance Systems 3961502, XPert Filtered Balance Systems 3961501, XPert Filtered Balance Systems 3961500, XPert Filtered Balance Systems 3950422, XPert Filtered Balance Systems 3950421, XPert Filtered Balance Systems 3950420, XPert Filtered Balance Systems 3950402, XPert Filtered Balance Systems 3950401, XPert Filtered Balance Systems 3950400, XPert Filtered Balance Systems 3950322, XPert Filtered Balance Systems 3950321, XPert Filtered Balance Systems 3950320, XPert Filtered Balance Systems 3950302, XPert Filtered Balance Systems 3950301, XPert Filtered Balance Systems 3950300, XPert Filtered Balance Systems 3950222, XPert Filtered Balance Systems 3950221, XPert Filtered Balance Systems 3950220, XPert Filtered Balance Systems 3950202, XPert Filtered Balance Systems 3950201, XPert Filtered Balance Systems 3950200