Chapter 6: Maintaining Your Filtered Enclosure
Product Service 1-800-522-7658
52
Figure 6-4
Motorized Impeller Replacement
Speed Control Replacement
XPert Filtered Balance System
Do NOT contact blower wheel while still in motion.
NE PAS être en contact avec la roué du ventilateur tant qu’il est
en marche.
1. Remove the bracket that the speed control is attached to. See Figures 6-1, 6-2
and 6-3.
2. Remove the two screws holding the speed control using a Phillips screwdriver.
Refer to Appendix A for Replacement Parts Diagram.
3. Disconnect all wires leading to the speed control. Connect wires on new speed
control in the same position as the old speed control.
4. Reassemble to the system in the same position and with the same screws that
were removed earlier.
!