Chapter 6: Maintaining the Cabinet

Product Service 1-800-522-7658

28

Service Operations

Door Evidence Prefilter Replacement:

1. Pull the prefilter frame(s) straight away from the door(s), as shown in

Figure 6-1.

2. Wearing appropriate gloves, and working on the inside surface of the

prefilter frame(s), carefully pull out the Evidence prefilter(s). The
Evidence prefilter(s) can now be examined for trace evidence, or stored
with the evidence in a sealed bag for future reference.

3. Examine the door prefilter(s). If it is contaminated from the evidence

drying operation, or from dust, replace it with a new one, such that the
orange side of the prefilter(s) faces you (away from the front of the
prefilter frame).

4. Press the prefilter frame(s) back into the door, engaging the fastener

strips.


Door Prefilter Replacement:

1. Follow Steps 1, 3, and 4 above.

Figure 6-1