Refer to the following table if the cabinet fails to operate properly. If the
suggested corrective actions do not solve the problem, contact Labconco for
additional assistance.

PROBLEM

CAUSE

CORRECTIVE ACTION

Cabinet blower and
lights won’t turn on

Unit not plugged into
outlet

Plug the cabinet into appropriate
electrical service.

Check connection to the top of cabinet.

Circuit

breaker(s)

tripped

Reset circuit breakers.

UV light not working

Door(s) are open

Close door(s) – UV light will not work
with the door(s) open.

UV

Security

Lock

not on

The key switch for the UV Security
Lock must be turned on.

Lamp is defective

Replace defective lamp.

Lamp wiring is
disconnected

Inspect lamp wiring.

Defective

lamp

ballast

Replace lamp ballast.

UV light is dim or
flickering

Lamp is defective or
is at end of operating
lifetime

Replace defective or worn out lamp.

Defective lamp
ballast

Replace lamp ballast.