Starting the freeze dry cycle – Labconco FreeZone Triad Freeze Dry System 74000 Series User Manual

Page 25

Advertising
background image

Chapter 4: Operating Instructions

Product Service 816-333-8811 or 1-800-522-7658

21

The display will show the parameters that were last entered for this program.






If any of the parameters are changed, the ENTER button must be pressed for
the new value to be stored. If ENTER is not pressed within 10 seconds after
the last parameter adjustment, the value will return to the previously stored
value. The flashing parameter name shows the parameter that will be
changed. Pressing ENTER only will advance to the next parameter.

After the AUTO RUN cycle is started, segments with a value greater than the
segment running may have their parameters changed. Only the selected
program may be modified. To modify a different program, the system must
be stopped and the desired program selected.

Starting the Freeze Dry Cycle

After the desired cycles have been programmed, push the MODE button to
select AUTO. Push the RUN/STOP button to start the program running. The
RUN/STOP indicator will light. The system display will change and the
display will change to MONITOR mode.

During the pre-freeze segment the status display will be as follows:

P1

Selected Program

SHLF

Shelf Temperature

SAMP

Sample Temperature (if probe is not
installed, the display shows ---)

°F or °C

Temperature Units

COL

No temperature displayed

SEG

Pre-Freeze Mode R(Ramp) H(Hold)

VAC

System Vacuum Level is above 5 mBar


During the ramp cycle of the active segment, the status display will be as
follows:

P1

Selected Program

SHLF

Shelf Temperature

SAMP

Sample Temperature (if probe is not
installed, the display shows ---)

°F or °C

Temperature Units

COL

Collector Temperature

SEG

Shows what segment is active and
Ramping (R)

VAC

System vacuum level

M

MBar

P

Pascal

T

Torr

P1 SEG 1 RAMP 00.0°C/MN
HOLD –34°C TIME 00.0

P1 SHLF = -70

SAMP= -70°C

COL = --- SEG = PR VAC = hi

P1 SHLF = -55

SAMP= -55°C

COL = -85 SEG = 1R VAC = .03M