Chapter 7: Troubleshooting

Product Service 816-333-8811 or 1-800-522-7658

39

Refrigeration Module Operation

In the Pre-freeze mode under a no-load condition and an ambient temperature
of 20°C, the freeze dry system refrigeration system can achieve a shelf
temperature of –75°C or lower within 4 hours.

If any repairs are required on the refrigeration module, please call Labconco.
Repairs should only be undertaken by a competent refrigeration technician or
through an authorized Labconco service agency.

Filling the Fluid System
































Figure 6