Chapter 8: Modifying Your Freeze Dry System Display

Product Service 816-333-8811 or 1-800-522-7658

43


If you do not want to restore all factory settings, but want to adjust the
vacuum offset, press DISPLAY briefly and the display will show:



Press  or  until the appropriate offset is shown.
Each number represents an offset of 0.001 mBar

Press DISPLAY and the display will show:


Press  or  until the appropriate offset is shown.
Each number represents an offset of 1°C.

Press DISPLAY and the display will show:


Press  or  until the appropriate offset is shown.
Each number represents an offset of 1°C.

Press DISPLAY and the display will show:


Press  or  until the appropriate offset is
shown. Each number represents an offset of 1°C.

Press DISPLAY and the display will show:



Press  or  until the appropriate offset is shown.
Each number represents an offset of 1°C.

Press DISPLAY and the display will show:





Wait approximately 10 seconds and the display will default to show
operational parameters.

VACUUM OFFSET: 0 mBar

PRODUCTION MENU: MASTER RESET?

COLLECTOR TEMP. OFFSET: 0°C

SHELF TEMP. OFFSET: 0°C

This is the probe mounted in the
shelf. Temperature from this probe
is displayed during Pre Freeze.

LEXSOL TEMP. OFFSET: 0°C

This is the probe mounted to the Lexsol tube
in the back. Temperature from this probe is
displayed as “SHLF TEMP” during
sequences other than PRE FREEZE.

SAMPLE TEMP. OFFSET: 0°C