Product Service 816-333-8811 or 1-800-522-7658

38

The shelf valve opens and refrigeration system turns on or press .

The display will show:



The collector valve opens. The (shelf valve closes) refrigeration system
remains on or press .

The display will show:


The refrigeration system remains on. The Lexsol pump turns on or press .

The display will show:


The defrost valve opens or press .

The display will show:


The refrigeration system turns off. The vacuum pump turns on or press .

The display will show:


The vacuum pump turns off. The vacuum control valve closes or press .

The display will show:




The vacuum control valve opens. The shelf tilt valve turns on (open) or press
.

The display will show:




The heater turns on for 3 minutes or press .