Before you install your Triad Freeze Dry System, you must be certain that the
area is level and of solid construction. An electrical source must be located
near the installation site.

Carefully read this chapter to learn:

Electrical supply requirements.

Location requirements.

Vacuum pump requirements.

Refer to Appendix C: Freeze Dry System Specifications for complete
electrical and environmental conditions, specifications and requirements.

Electrical Requirements

The Triad Freeze Dry System requires a dedicated electrical outlet. This
outlet requires a 20 Amp circuit breaker or fuse for all models. Various
power cord configurations are provided with 230V models. If this does not
match with the available receptacle, remove this plug and replace it with an
approved plug of the suitable style.

Location Requirements

The Triad Freeze Dry System should be located in an area that provides an
unobstructed flow of air around the cabinet. This air cools the refrigeration
system. The refrigeration system draws air in through the left side and
exhausts it through the right side. A minimum of 3" must be allowed
between the rear and both sides of the Triad Freeze Dry System and adjacent
wall surfaces. Restriction of airflow during operation could adversely affect
performance. The Triad Freeze Dry System must be placed on a work surface
or table that will support over 400 lbs.

Refer to Appendix B: Freeze Dry System Dimensions for drawings.