Troubleshooting – Nexen RSTC1100 964528 User Manual

Page 25

Advertising
background image

25

FORM NO. L-21204-C-0705

TROUBLESHOOTING

Problem

Probable Cause

Diagnosis Test

Corrective Action

Power Indicator is not
illuminated.

No power

Check for 24 VDC across

terminals 1 and 2 using a
voltmeter.

Restore 24 VDC.

Control inputs do not
respond.

Improper or missing signal

Measure the voltage at inputs
41, 43 and 44.

Correct any problems with

voltage or wiring using the
SPECIFICATIONS section to
determine proper range.

RSTC will not go
into Auto mode.

Run signal not provided to

control input or software
command not given

Start machine and watch for

Run/Stop to go high on
Diagnostics window.

Check Run/Stop input wiring or
control panel to make sure the
command is sent properly.

Run/Stop signal not set up
properly

Check Run/Stop configuration
in Signal Calibration window.

Refer to Signal Calibration

section of manual and select

proper configuration.

RSTC in manual mode

Check Automatic/Manual
button on Diagnostics window.

Switch to Automatic using
Automatic/Manual button or

issue Automatic command from

control panel.

Tension fluctuates in
automatic mode.

Improper settings or

calibration

Tension not fluctuating in

manual mode indicates a
problem with RSTC.

Re-tune RSTC.

Recalibrate load cells.

Web machine problem

Fluctuating tension in manual
mode indicates a machine
problem.

Correct problem with web
machine.

Output goes to 0% during
unwinding roll
acceleration.

Machine acceleration rate too
high

Decrease acceleration rate and

see if output % increases
above 0%.

Decrease machine acceleration

as roll inertia limits acceleration

rate of machine.

Output goes to 0% and

tension rises above
setpoint when roll unwinds
at constant web speed.

Acceleration of the unwinding

roll shaft causing tension
increase

Decrease web speed as roll

diameter approaches core to
see if output increases above 0.

Decrease web speed as roll

diameter approaches core
because roll and shaft inertia
cause tension to increase
above setpoint.

Output goes to 100%
during deceleration and

tension falls below
setpoint.

Machine deceleration rate is

too high and there is not
enough torque.

Decrease deceleration rate and

see if output stays below

100%.

Decrease machine deceleration

as roll inertia limits deceleration

rate of machine or increase

torque output of torque
actuator.

Output goes to 100% and

tension falls below
setpoint when roll winds
at constant web speed.

Not enough torque for the size

of roll and the tension setpoint

Decrease the tension setpoint
(increase taper) and see if

output stays below 100%.

Increase torque output of

torque actuator or decrease
tension setpoint or increase
taper.

Advertising
This manual is related to the following products: