Fl {_h_l {hi, Fl {_h_l {s_m, M_n {b;l {nb – CIRCUTOR CVM-MINI Series User Manual

Page 15: M_n {b;l

Advertising
background image

Page 15 of 26

4.1.4.3

Backlight (Backlit display)

The time in which the back lighting will be on after the last use is set by using
the keypad in this menu. The display is permanently on if

** is set.

4.1.5 Clearing energy meter values

Clearing energy meters refers to the four consumed or imported energy meters.
To select clear these meters of kw·h, press the

max

key and the two options will

alternate. Once the required option is selected, press the

key to enter the

data and access the next setting step.

{]fl

{_h_l

{hi

{]fl

{_h_l

{s_m

Figure 44. No-clearing energy meters

Figure 45. Clearing energy meters

4.1.6 Setting THd or d

Harmonic distortion rate measurements may be made using two measurement
options: the effective value or the fundamental value.
To select one of the two calculation options, press the

max

key and the two

options will alternate. Once the required option is selected, press the

key to

enter the data and access the next setting step.

{m_n

{B;L

{NB>

{m_n

{B;L

{>

Figure 46. Harmonic distortion value for the

effective value (RMS)

Figure 47. Harmonic distortion value for the

fundamental value

4.1.7 Digital output for the transistor (2)

Two types of setting may be made using the CVM-MINI’s digital transistor
outputs:

• Impulse per n kW·h or kvar·h (Energy): the value for the energy consumed

or generated may be set to generate an impulse.

• Alarm condition: associates a parameter to a digital output, setting a

maximum, minimum and delay (delay) for the trip condition.

In the event that no condition needs to be set, the

** code is typed and entered

using the

key.

Advertising