AMETEK RS Series User Manual
Page 153

User Manual
AMETEK Programmable Power
RS Series
153
6.4.5 Output Calibration Summary
The following Table is a summary of the preceding calibration steps. Program the following
values in the table and make the adjustments in the OUTPUT CALIBRATION screen. Select the
phase to be calibrated by pressing the PHASE key.
CALIBRATION
PROGRAM VALUES
CALIBRATION VALUE
ADJUST TO
300 VAC range DC Zero
300 VAC range, 0.0 V,
60 Hz
VOLT ZERO
0 ± 5 mV DC
300 VAC range High Freq
DC Zero, -HF models only.
300 VAC range, 0.0 V,
820 Hz
VOLT ZERO
0 ± 5 mV DC
300 VAC range Volt FS
240.0 V, 60 Hz
VOLT FS
240
±
0.05 VAC
400 VDC range DC Zero
400 VDC range, 0.0 V
VOLT ZERO
0 ± 5 mV DC
Phase Offset B, C
150 VAC range, 120V,
60 Hz
PHASE OFST
± 0.5
°
Option -413
400 Hz, 20VAC
1800 Hz, 20VAC
R9, R10, R11
INTER HARM FS
20 ± 1.0 Vrms
20 ± 1.0 Vrms
Table 6-3: Output Calibration Table – RS Series