Number
Message String
Cause
Remedy
too high and could damage
amplifier.
programmed.
23
"ALC or Impedance
must be off"
Conflict between ALC and
programmable impedance
mode.
Turn off ALC to use
programmable impedance. Turn
off programmable impedance to
use ALC.
29
“DC component
exceeds limit”
The waveform selected
contains a DC offset that
exceeds the AC mode
capability.
Select AC+DC mode.
30
“Amplifier fault”
Amplifier fault.
Contact customer service.
31
“Warning negative
power near li
mit”
Approaching limit on the
amount of power that can
be fed back into the supply
by an active load. This is a
warning only.
Stop increasing power feedback
into the power supply. Typically
occurs when using AC
inverters. If power increases
further, an error 32 will be
generated.
32
“Negative power fault”
Too much power fed back.
Power source output
disconnected.
Reduce the amount of power
being fed back into the power
source.
Table 10-1: Error Messages