Table 2-5. channel a dvm calibration settings -6 – AMETEK SL Series Software User Manual
Page 14

Calibration Procedure
Elgar Electronics Corporation
DVM, Channel A
1.
Set the DC Voltage Standard to the value shown in DC Input column
of Table 2-5.
2.
Adjust using
keys until the DMM reading matches the value in
DMM Reading column.
3.
Scroll to next item (press
) and repeat Steps 1 and 2 for each item
through Item 8.
4.
Repeat Steps 1 through 3 for all calibration parameters, Items 5
through 8 (press
three times to return to Item 5) until no further
adjustments are necessary for the DMM reading to match the value in
the DMM Reading column of Table 2-5.
ITEM
DC INPUT
DMM Reading
5
0.000 V
0.000 V
6
15.0000 V
15.000 V
7
0.000 V
0.000 V
8
60.000 V
60.00 V
*
80.000 V
80.00 V
*
Applies to SLD-80-20-102 only
Table 2-5. Channel A DVM Calibration Settings
5.
Scroll to next calibration parameter, Item 9 (press
).
2-6
SLD Calibration Manual