Table 2-16. channel b dvm calibration settings -18 – AMETEK SL Series Software User Manual
Page 26

Calibration Procedure
Elgar Electronics Corporation
DVM, Channel B
1.
Set the DC Voltage Standard to the values shown in DC Input column
of Table 2-16.
2.
Adjust using
keys until the DMM reading matches the value in
DMM Reading column.
3.
Scroll to next item (press
) and repeat Steps 1 and 2 for each item
through Item 8.
4.
Repeat Steps 1 through 3 for all calibration parameters, Items 5
through 8 (press
three times to return to Item 5) until no further
adjustments are necessary for the DMM reading to match the value in
the DMM Reading column of Table 2-16.
ITEM
DC INPUT
DMM Reading
5
0.000 V
0.000 V
6
15.0000 V
15.000 V
7
0.000 V
0.000 V
8
60.000 V
60.000 V
*
80.000 V
80.000 V
*
Applies to SLD-80-20-102 only
Table 2-16. Channel B DVM Calibration Settings
5.
Scroll to next calibration parameter, Item 9 (press
).
2-18
SLD Calibration Manual